Internal vs. External Excitation: Configuring the Pickering PXI/PXIe LVDT/RVDT/Resolver Simulator Modules

The Pickering LVDT/RVDT/Resolver Simulator Modules [41–670 (PXI) and 43–670 (PXIe)] are purpose-built for hardware-in-the-loop (HIL) test environments where physical transducers can't be used directly. They let you replicate LVDT, RVDT, and resolver output signals electronically, without requiring live hardware or physical displacement, making them well suited to validating closed-loop control systems in aerospace, automotive, and industrial applications.

One configuration decision shapes signal integrity, transfer function behavior, and how closely your test environment mirrors the real-world system: whether to use internal or external excitation. This guide covers the physics behind each approach, how to configure them on the 41–670 and 43–670, and how to select the right method for your unit under test (UUT) and test objectives.

Module Overview

Each module provides up to four banks. Each bank simulates a single 5- or 6-wire LVDT, RVDT, or resolver, or two independent 4-wire transducers sharing a common excitation. Inputs and outputs are galvanically isolated via transformers to closely emulate real-world transducer behavior. Onboard fault-insertion relays simulate open- and short-circuit conditions, supporting fault-coverage validation in safety-critical systems.

The modules support excitation frequencies from 300 Hz to 20 kHz, input voltages up to 38 V, and output voltages up to 31 V.

Internal Excitation

With internal excitation, our LVDT/RVDT/Resolver Simulator Modules generate their own primary excitation signal and use it to derive the secondary winding outputs. No external signal generator is required.

How It Works

Each bank's internal excitation circuitry produces a sine wave at a user-configured frequency within the 300 Hz to 20 kHz operating range. The module then produces secondary outputs modulated according to the configured displacement or angular position. The phase relationship between the excitation and the secondary outputs is automatically adjusted to lag by one cycle. Where an in-phase output is needed for demodulation, one output can be configured to propagate the input signal directly.

Because the excitation source and simulation logic share the same module, the transfer function from excitation to secondary outputs is tightly controlled. This minimizes the risk of phase mismatch or amplitude variation due to external cabling or instrument drift, making internal excitation well-suited to applications where repeatability and signal-path simplicity take priority.

Configuration Steps

  1. Select Internal under the Input section of the module's control panel.
  2. Set the excitation frequency for your application.
  3. Configure output amplitude using Vsum/Vdiff (for 5- or 6-wire configurations), percentage displacement, or independent voltage outputs.
  4. Set any required phase delay or actuator response rate.
  5. Click Sample Load to apply settings.
  6. Close the OUTA and OUTB switches to enable output.
  7. Confirm secondary outputs on the oscilloscope before proceeding to automated test sequences.

When to Use Internal Excitation

  • Validating controller logic or signal processing algorithms independent of a specific excitation source.
  • Early-stage bring-up or algorithm verification where excitation profile fidelity isn't yet a constraint.
  • Minimizing setup time and reducing the number of instruments in the signal path.
LVDT Internal Excitation How-To Video

External Excitation

External excitation uses a separate signal source to supply the module's primary excitation. The LVDT/RVDT/Resolver Simulator Modules accept independent excitation inputs per bank and support a single shared excitation distributed across all banks, which reduces cabling complexity in multi-channel test rigs.

How It Works

An external signal generator outputs a sine wave at the target excitation frequency. The module reads the incoming frequency and amplitude, then produces secondary outputs modulated according to the configured position or angle. An auto input-gain routine calibrates the module to the incoming signal level before enabling the output.

The programmable phase delay allows you to offset the output relative to the excitation, simulating imperfect sensors or non-ideal cabling behavior. This is particularly useful when verifying how a signal conditioner or controller handles phase variation. Where synchronization with other HIL components is required, external excitation also lets you tie the simulation reference to a common source, maintaining phase coherence across the test environment.

Configuration Steps

  1. Connect the external signal source to the bank's excitation input.
  2. Close the IN switches.
  3. Run Auto Input Gain to calibrate the module to the incoming signal amplitude. Skipping this step is one of the most common causes of signal distortion in external excitation setups.
  4. Verify excitation frequency and amplitude readback in the control interface.
  5. Configure output amplitude and phase parameters.
  6. Click Sample Load to apply settings.
  7. Close the OUTA and OUTB switches to enable output.
  8. Confirm secondary outputs on the oscilloscope before beginning automated sequences.

When to Use External Excitation

  • Your test requires the module to respond to the excitation produced by the actual signal conditioner or controller under test.
  • You need excitation waveform control beyond the module's internal generation capability.
  • You're validating system-level behavior where excitation synchronization across HIL simulation channels is required.
  • You need to simulate sensor degradation or cabling faults through controlled phase delay or amplitude offsets.
lvdt-external-excitation-how-to

Monitoring and Troubleshooting

Use a calibrated oscilloscope to monitor secondary outputs throughout configuration: sine output on Channel 0 and cosine output on Channel 1 for resolver simulation. Always confirm signal presence and waveform shape before running automated test sequences.

Common issues and resolutions:

  • No output signal: Confirm OUTA and OUTB switches are closed and the Input mode matches your intended excitation source.
  • Distorted or clipped output (external mode): Re-run Auto Input Gain. A mismatch between the incoming signal level and the module's gain setting is the most frequent cause.
  • Output drops unexpectedly (external mode): Opening the IN switches removes the excitation reference and disables output. Check the IN switch state first.
  • Residual output after disabling the external generator: Disabling the signal source and closing the module's output switches are independent actions. Both must be addressed to fully de-energize the simulation output.
  • Unacceptable phase lag in secondary outputs: Configure one output to propagate the input signal directly for use as a demodulation reference.

Selecting the Right Excitation Method

Consideration Internal Excitation External Excitation
Test objective Validating controller logic independent of excitation source Replicating a specific real-world excitation profile
System complexity Simpler setup, fewer instruments Requires external signal source and calibration
Synchronization needs Minimal, self-contained Required when matching other HIL simulation channels
Instrument availability No additional generator needed Requires a compatible external signal source
Signal path predictability High, contained within the module Dependent on calibration and cabling quality
Fault simulation capability Standard fault insertion via onboard relays Extended fault simulation via phase delay and amplitude control

Best practices

  • Use internal excitation for initial bring-up and controller logic validation where matching a specific excitation profile isn't yet required.
  • Switch to external excitation when system-level validation requires the module to respond to the actual excitation from the controller or signal conditioner under test.
  • Always run Auto Input Gain after connecting an external excitation source, before enabling output switches.
  • Use the module's programmable phase delay to simulate sensor aging, cabling imperfections, or controlled phase offsets as part of fault coverage testing.
  • Document excitation configuration—frequency, amplitude, gain settings, phase delay—alongside test results to ensure reproducibility across test runs and team members.

To discuss configuration options or to specify a module for your application, contact Pickering's applications engineering team.

Frequently Asked Questions

Can I switch between internal and external excitation without reconfiguring the full bank?

Yes. The input source selection is a discrete configuration parameter. You can switch between Internal and External modes without changing output amplitude or position settings. When switching to external excitation, always re-run Auto Input Gain before enabling output.

What happens if my external excitation frequency drifts during a test?

Our LVDT/RVDT/Resolver Simulator Modules include per-channel readbacks of excitation frequency and amplitude. If the incoming excitation frequency changes, the module's output will track it, but phase and amplitude relationships may shift. Monitor the excitation readback during test execution and re-run Auto Input Gain if a significant drift occurs.

Can I share a single external excitation source across multiple banks?

Yes. The modules support single-input distribution across all banks, reducing the number of required excitation sources and simplifying cabling in multi-channel test rigs.

Can these modules integrate with other Pickering simulation hardware in the same chassis?

Yes. Our LVDT/RVDT/Resolver Simulator Modules are part of Pickering's broader PXI and PXIe sensor-simulation platform, which includes RTD, thermocouple, strain-gauge, and 4–20 mA current-loop simulators. All modules are compliant with PXI standards and can be housed in any compliant chassis.

What's the maximum resolver rotation speed these modules can simulate?

Certain configurations of the 41–670 and 43–670, specifically the high-speed resolver variants, support rotation simulation up to 130,000 RPM.

How did we do?
0 out of 0 people found this helpful.
Beta

How can I help?

Hi , I'm TESS - your friendly AI agent

Press Enter to submit

TESS is a moderated, AI agent. TESS can make mistakes, please verify important information independently.


Chat history

Log in or create an account to save & continue your chats with TESS.

Log in

or create an account